FDA Adverse Event Injury Summary report: N

PUMP MMT-722NAS PRDGM INS V2.2 SK EN

MDR report key: 1931205 · Received December 10, 2010

Report

Report Number
2032227-2010-83400
Event Type
Injury
Date Received
December 10, 2010
Date of Event
November 21, 2010
Report Date
November 23, 2010
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT SHE WAS HOSPITALIZED FOR DIABETIC KETOACIDOSIS, WITH A BLOOD GLUCOSE READING OF 608 MG/DL. THE CUSTOMER STATED THAT THE CANNULA WAS BENT WHEN SHE REMOVED THE INFUSION SET, BUT SHE NEVER GOT A NO DELIVERY ALARM. THE CUSTOMER CHANGED THE INFUSION SET, AND THAT CANNULA WAS BENT ALSO. FOUND THAT THE CUSTOMER HAS LOST BODY FAT AND HAS BEEN EXERCISING MORE. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP PASSED THE PRIME AND HIGH PRESSURE TESTS. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAS PRDGM INS V2.2 SK EN INFUSION PUMP MDS MEDTRONIC MINIMED MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 33 YR Hospitalization