FDA Adverse Event Malfunction Summary report: N

DELTA

MDR report key: 19311689 · Received May 14, 2024

Report

Report Number
1220063-2024-00056
Event Type
Malfunction
Date Received
May 14, 2024
Date of Event
April 30, 2024
Report Date
September 5, 2024
Manufacturer
DRAEGER INC.
Product Code
MHX
UDI-DI
04049098004152
PMA / PMN Number
K152407
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION HAS JUST STARTED; RESULTS WILL BE PROVIDED IN A FOLLOW-UP REPORT.

Additional Manufacturer Narrative · 0

IT WAS REPORTED TO DRAEGER THAT THE DELTA POWER SUPPLY BECAME CHARRED AND CAUGHT FIRE WHEN IT WAS PLUGGED INTO THE MAIN AC POWER SOURCE. VISUAL INSPECTION OF THE POWER SUPPLY CONFIRMED THAT A THERMAL EVENT HAD OCCURRED, CAUSED BY THE ENERGY RELEASE DURING A SHORT CIRCUIT. THE DRÄGER SERVICE ENGINEER AT THE CUSTOMER'S SITE REPORTED THAT THE CITED POWER SUPPLY WAS MOUNTED ON A TRANSPORT TROLLEY WITH THE AC MAIN POWER CONNECTION FACING UPWARDS. IN AN INTERVIEW WITH THE BIOMEDICAL ENGINEER IN THIS WARD, BOTH SIDES CONCLUDED THAT DISINFECTANT LIQUID HAD PENETRATED THE (AC MAIN POWER) PLUG CONNECTION AND CAUSED A SHORT CIRCUIT WHICH DAMAGED THE DEVICE. IT COULD BE CONFIRMED THAT THE DURATION OF THE EVENT WAS ONLY 1 OR 2 SECONDS UNTIL THE FUSE TRIPPED. THE TROLLEY INSTALLATION WASN¿T MADE BY DRÄGER. A NEW POWER SUPPLY WAS HANDED OVER TO THE CUSTOMER. DRÄGER FINALLY CONCLUDES THAT THE UNFAVORABLE WAY OF MOUNTING OF THE EQUIPMENT ON THE TROLLEY CONTRIBUTED TO THE EVENT. MOST LIKELY, THE PRESENCE OF DISINFECTANT LIQUID HAS CAUSED A SHORT CIRCUIT OR DRIED RESIDUES OF IT HAD FORMED A CREEPING PATH WITH THE SAME CONSEQUENCE. THE DELTA MONITOR AND THE ASSOCIATED UNIVERSAL POWER SUPPLY IS DESIGNED TO MEET THE IEC 60601-1 STANDARD FOR BASIC SAFETY AND ESSENTIAL PERFORMANCE OF MEDICAL ELECTRICAL EQUIPMENT AND THE RISK OF FIRE IS MINIMIZED BECAUSE THE USED MATERIALS ARE CLASSIFIED AS UL 94 V-0 OR BETTER ¿ THE IGNITION OF THE MATERIAL WILL EXTINGUISH BY ITSELF AS SOON AS THE ENERGY SUPPLY IS CUT OFF (TRIPPING FUSE OPENS THE ELECTRIC CIRCUIT). THERE IS NO OTHER SIMILAR EVENT KNOWN.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT THE DELTA POWER SUPPLY UNIT CAUGHT FIRE WHEN PLUGGED INTO THE MAIN POWER SUPPLY. THE POWER SUPPLY UNIT AND DOCKING STATION WERE MOUNTED ON A TROLLEY, WITH THE MAINS PLUG FACING UPWARDS. THE TRANSPORT TROLLEY WAS CONNECTED TO A PATIENT AT THE TIME OF THE INCIDENT, AND THE DELTA WAS CONNECTED TO THE POWER SUPPLY TO BRIDGE A WAITING PERIOD. THE PLUG OF THE POWER SUPPLY UNIT WAS CONNECTED TO A MULTIPLE SOCKET ON THE TROLLEY, AND THERE WAS A BANG AND A "JET OF FLAME" IN THE AREA OF THE POWER SUPPLY UNIT. NO PATIENT OR USER INJURY OR DEATH WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT THE DELTA POWER SUPPLY UNIT CAUGHT FIRE WHEN PLUGGED INTO THE MAIN POWER SUPPLY. THE POWER SUPPLY UNIT AND DOCKING STATION WERE MOUNTED ON A TROLLEY, WITH THE MAINS PLUG FACING UPWARDS. THE TRANSPORT TROLLEY WAS CONNECTED TO A PATIENT AT THE TIME OF THE INCIDENT, AND THE DELTA WAS CONNECTED TO THE POWER SUPPLY TO BRIDGE A WAITING PERIOD. THE PLUG OF THE POWER SUPPLY UNIT WAS CONNECTED TO A MULTIPLE SOCKET ON THE TROLLEY, AND THERE WAS A BANG AND A "JET OF FLAME" IN THE AREA OF THE POWER SUPPLY UNIT. NO PATIENT OR USER INJURY OR DEATH WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1102570 DELTA PHYSIOLOGIC MONITORING SYSTEM MHX DRAEGER INC. 04049098004152

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other N/A.| N/A.