FDA Adverse Event Malfunction Summary report: N

GEMSTR 7 THPY ALT ST

MDR report key: 1931164 · Received November 11, 2010

Report

Report Number
2921482-2010-00877
Event Type
Malfunction
Date Received
November 11, 2010
Date of Event
October 1, 2010
Report Date
October 12, 2010
Manufacturer
HOSPIRA, INC.
Product Code
FRN
PMA / PMN Number
K042980
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THE DELIVERY DID NOT STOP WHEN THE STOP KEY ON THE DEVICE KEYPAD WAS PRESSED. THIS WAS DUE TO A BROKEN KEYPAD. THE CAUSE OF WAS WEAR AND TEAR OF THE KEYPAD IN THE CUSTOMER ENVIRONMENT. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL. (B)(4)

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE DID NOT STOP DELIVERING WHEN THE STOP KEY WAS PRESSED. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GEMSTR 7 THPY ALT ST 80FRN FRN HOSPIRA, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA