FDA Adverse Event Injury Summary report: N

NOVASURE IMPEDANCE CONTROLLED EA SYSTEM

MDR report key: 1931147 · Received December 9, 2010

Report

Report Number
1222780-2010-00190
Event Type
Injury
Date Received
December 9, 2010
Date of Event
November 9, 2010
Report Date
November 9, 2010
Manufacturer
HOLOGIC
Product Code
MNB
PMA / PMN Number
P010013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IS NOT BEING RETURNED; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE CAN NOT BE COMPLETED. BASED ON THE INFORMATION OBTAINED TO DATE, NO DIRECT CORRELATION CAN BE MADE BETWEEN THE REPORTED EVENT AND THE NOVASURE SYSTEM. DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR THE REPORTED LOT NUMBER. THE LOT WAS RELEASED MEETING ALL QA SPECIFICATIONS. CURRENTLY UNABLE TO ESTABLISH A RELATIONSHIP OR IMPACT TO THE REPORTED OBSERVATION. ACCORDING TO THE INSTRUCTIONS FOR USE (IFU) WARNINGS: USE CAUTION NOT TO PERFORATE THE UTERINE WALL WHEN SOUNDING, DILATING, OR INSERTING THE DISPOSABLE DEVICE. IF THE DISPOSABLE DEVICE IS DIFFICULT TO INSERT INTO THE CERVICAL CANAL, USE CLINICAL JUDGMENT TO DETERMINE WHETHER OR NOT FURTHER DILATION IS REQUIRED. THE NOVASURE SYSTEM PERFORMS A CAVITY INTEGRITY ASSESSMENT (CIA) TEST TO EVALUATE THE INTEGRITY OF THE UTERINE CAVITY, AND SOUNDS AN ALARM WARNING OF A POSSIBLE PERFORATION PRIOR TO TREATMENT. (B)(4).

Description of Event or Problem · 1

THREE UNSUCCESSFUL CAVITY INTEGRITY ASSESSMENT (CIA) TESTS OCCURRED DURING AN ATTEMPTED NOVASURE ENDOMETRIAL ABLATION. THE PROCEDURE WAS ABANDONED AND A PERFORATION IN THE FUNDUS WAS CONFIRMED ON POST HYSTEROSCOPY. NO TREATMENT WAS NEEDED FOR THE PERFORATION AND THE PT WAS DISCHARGED HOME. ON (B)(6) 2010, THE PHYSICIAN REPORTED THE PT HAS BEEN SEEN ON F/U AND IS "FINE". A TUBAL LIGATION, HYSTEROSCOPY, DILATATION, AND SOUNDING (NOT A HOLOGIC DEVICE) WERE PERFORMED PRIOR TO THE ATTEMPTED ABLATION. IT IS NOT KNOWN WHEN THIS PERFORATION OCCURRED OR WHAT INSTRUMENT MAY HAVE BEEN THE CAUSE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NOVASURE IMPEDANCE CONTROLLED EA SYSTEM MNB HOLOGIC NS2000 10D13RA

Patients

Seq Age Sex Outcome Treatment
1 UNK Other RADIO FREQUENCY CONTROLLER: SERIAL # UNK