FDA Adverse Event Malfunction Summary report: N

HEARTMATE II SYSTEM CONTROLLER

MDR report key: 1931115 · Received November 10, 2010

Report

Report Number
2916596-2010-00295
Event Type
Malfunction
Date Received
November 10, 2010
Date of Event
September 27, 2010
Report Date
October 12, 2010
Manufacturer
THORATEC CORP.
Product Code
DSQ
Removal / Correction Number
29165969/2/10001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SYSTEM CONTROLLER WAS RETURNED TO THE MFR FOR EVALUATION AND THE INTERMITTENT ALARMS WERE CONFIRMED DURING THE ANALYSIS OF THE CONTROLLER. DURING FUNCTIONAL TESTING OF THE CONTROLLER, THE "POWER CABLE DISCONNECT" ALARMS OCCURRED WHEN THE BLACK POWER LEAD WAS MANEUVERED AT THE CONNECTOR END OF THE LEAD. FURTHER EVALUATION OF THE SYSTEM CONTROLLER REVEALED THAT THE BROWN CONDUCTOR IN THE BLACK POWER LEAD WAS BROKEN. THE LOG FILE WAS ALSO REVIEWED AND SUPPORTED THE REPORTED EVENT. THIS SITUATION IS BEING ADDRESSED THROUGH THE MFR'S CORRECTIVE/PREVENTATIVE ACTION SYSTEM AND AN URGENT MEDICAL DEVICE CORRECTION NOTICE (29165969/2/10001-C) WAS SENT TO CUSTOMERS. NO FURTHER INFO IS AVAILABLE. THE MFR IS CLOSING ITS FILE ON THIS EVENT.

Description of Event or Problem · 1

THE PT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE (LVAD). IT WAS REPORTED BY THE VAD COORD THAT THE PT EXPERIENCED INTERMITTENT BEEPING WHEN HE MOVED OR WAS LYING STILL IN BED. THIS OCCURRED WHILE HE WAS CONNECTED TO THE POWER MODULE OR ON BATTERIES. A FEW WEEKS PRIOR, THIS ISSUE HAD OCCURRED AND THE PT'S BATTERY CLIPS WERE EXCHANGED AND HIS BATTERIES WERE CLEANED AND INSPECTED. THIS TIME THE VAD COORD EXCHANGED THE SYSTEM CONTROLLER WITHOUT INCIDENT. THE PT REMAINS ONGOING ON THE LVAD AND NO FURTHER ISSUES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HEARTMATE II SYSTEM CONTROLLER LVAD SYSTEM CONTROLLER DSQ THORATEC CORP. 103696 NA

Patients

Seq Age Sex Outcome Treatment
1 60 YR Other