HEARTMATE II SYSTEM CONTROLLER
Report
- Report Number
- 2916596-2010-00295
- Event Type
- Malfunction
- Date Received
- November 10, 2010
- Date of Event
- September 27, 2010
- Report Date
- October 12, 2010
- Manufacturer
- THORATEC CORP.
- Product Code
- DSQ
- Removal / Correction Number
- 29165969/2/10001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
THE SYSTEM CONTROLLER WAS RETURNED TO THE MFR FOR EVALUATION AND THE INTERMITTENT ALARMS WERE CONFIRMED DURING THE ANALYSIS OF THE CONTROLLER. DURING FUNCTIONAL TESTING OF THE CONTROLLER, THE "POWER CABLE DISCONNECT" ALARMS OCCURRED WHEN THE BLACK POWER LEAD WAS MANEUVERED AT THE CONNECTOR END OF THE LEAD. FURTHER EVALUATION OF THE SYSTEM CONTROLLER REVEALED THAT THE BROWN CONDUCTOR IN THE BLACK POWER LEAD WAS BROKEN. THE LOG FILE WAS ALSO REVIEWED AND SUPPORTED THE REPORTED EVENT. THIS SITUATION IS BEING ADDRESSED THROUGH THE MFR'S CORRECTIVE/PREVENTATIVE ACTION SYSTEM AND AN URGENT MEDICAL DEVICE CORRECTION NOTICE (29165969/2/10001-C) WAS SENT TO CUSTOMERS. NO FURTHER INFO IS AVAILABLE. THE MFR IS CLOSING ITS FILE ON THIS EVENT.
THE PT WAS IMPLANTED WITH A LEFT VENTRICULAR ASSIST DEVICE (LVAD). IT WAS REPORTED BY THE VAD COORD THAT THE PT EXPERIENCED INTERMITTENT BEEPING WHEN HE MOVED OR WAS LYING STILL IN BED. THIS OCCURRED WHILE HE WAS CONNECTED TO THE POWER MODULE OR ON BATTERIES. A FEW WEEKS PRIOR, THIS ISSUE HAD OCCURRED AND THE PT'S BATTERY CLIPS WERE EXCHANGED AND HIS BATTERIES WERE CLEANED AND INSPECTED. THIS TIME THE VAD COORD EXCHANGED THE SYSTEM CONTROLLER WITHOUT INCIDENT. THE PT REMAINS ONGOING ON THE LVAD AND NO FURTHER ISSUES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTMATE II SYSTEM CONTROLLER | LVAD SYSTEM CONTROLLER | DSQ | THORATEC CORP. | 103696 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Other |