FDA Adverse Event Malfunction Summary report: N

ENDOPATH** XCEL

MDR report key: 1931106 · Received December 17, 2010

Report

Report Number
3005075853-2010-07227
Event Type
Malfunction
Date Received
December 17, 2010
Date of Event
December 7, 2010
Report Date
December 8, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GCJ
PMA / PMN Number
K032676
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. UPON EVALUATION OF THE DEVICE, THE DUCKBILL WAS FOUND TO BE SLIGHTLY OPEN AT THE SLIT. A LEAK TEST WAS PERFORMED AND THE DEVICE WAS NOTED TO LEAK WITHOUT THE TEST PROBE INSERTED THROUGH THE DEVICE. A POTENTIAL CAUSE OF THIS FAILURE IS ATTRIBUTED TO MOLDING, COMPONENT TRANSIT, BODILY FLUIDS OR DEBRIS FROM SURGERY. AS EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING THE MANUFACTURING PROCESS, NO CONCLUSION COULD BE REACHED AS TO WHAT MIGHT HAVE CAUSED THE REPORTED EVENT. HOWEVER, AN INVESTIGATION HAS BEEN INITIATED TO ADDRESS THE POTENTIAL ROOT CAUSE OF THE INSUFFLATIONS ISSUES. IN ADDITION THE LENS OF THE OBTURATOR WAS NOTED MELTED. ONE POSSIBLE CAUSE FOR THIS TYPE OF DAMAGE MAY BE INTERACTION WITH AN ENERGIZED DEVICE USED DURING THE PROCEDURE. CAUTION SHOULD BE TAKEN TO AVOID CONTACT BETWEEN AN ENERGIZED DEVICE AND THE TROCAR DURING THE SURGICAL PROCEDURE. HOWEVER, THIS FINDING IS NOT RELATED WITH THE INCIDENT REPORTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC VENTRAL HERNIA PROCEDURE, THE FIRST DEVICE WAS LEAKING FROM THE SIDE OF THE CAP, POSSIBLY FROM TORQUING THE DEVICE. THEN IT WAS REPLACED AND IT WAS ALSO LEAKING BUT MANAGED THE LEAK AND COMPLETED THE CASE WITH THIS DEVICE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT. IT WAS UNKNOWN IF THE DEVICES WERE HISSING WHEN LEAKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH** XCEL LAPAROSCOPE, GENERAL AND PLASTIC SURGERY GCJ ETHICON ENDO-SURGERY, LLC. NA G4U191