CONCERTO CRT-D DR
Report
- Report Number
- 2647346-2010-00937
- Event Type
- Death
- Date Received
- December 17, 2010
- Date of Event
- November 17, 2010
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- NIK
- PMA / PMN Number
- P010031/S031
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- NURSE
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): NO ANOMALIES FOUND. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. (B)(4): A RESISTANCE INCREASE IS SEEN ON EACH LEAD BETWEEN (B)(6) 2010 AND (B)(6) 2010: FOR THE RV PACE LEAD IMPEDANCE, FROM 672 TO 824 OHMS; FOR THE LV PACE LEAD IMPEDANCE, FROM 592 TO 992 OHMS; FOR THE A PACE LEAD IMPEDANCE, FROM 496 TO 1048 OHMS, FOR THE DEFIB AND SVC DEFIB IMPEDANCES, 45 TO 60 OHMS, AND 56 TO 102 OHMS. OUT OF TOLERANCE LEAD IMPEDANCE ALERT IS RECORDED ON (B)(6) 2010 AT 02:15:05 FOR THE SVC LEAD IMPEDANCE AT 102 OHMS. THE HIGH ALERT SETTING IS 100 OHMS FOR THIS LEAD. 17 V-SIC WERE RECORDED BEGINNING (B)(6) 2010. THE LIFETIME COUNTER HAD ZERO COUNTS PRIOR TO THIS TIME. INCREASED SIC CAN INDICATE NOISE IN THE SYSTEM. TWO SHORT V-V SENSED EVENTS OF < 220 MS ARE OBSERVED ON THE (B)(6) 2010. (2 EPISODES NST). (B)(4): NO ANOMALIES FOUND, ALL CONDUCTORS STRETCHED, OUTER INSULATION COSMETIC DEPRESSION, LEAD STRETCHED, APPARENT EXPLANT DAMAGE. PROXIMAL SEGMENT RETURNED AND ANALYZED. (B)(4): NO ANOMALIES FOUND, OUTER INSULATION COSMETIC ESC. PROXIMAL SEGMENT RETURNED AND ANALYZED. (B)(4): NO ANOMALIES FOUND, INNER INSULATION PULLED APART (OVERSTRESS), INNER TUBING TORN, OUTER INSULATION COSMETIC DEPRESSION, APPARENT EXPLANT DAMAGE. PROXIMAL SEGMENT RETURNED AND ANALYZED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY: (B)(4): NO ANOMALIES FOUND, ALL CONDUCTORS STRETCHED, OUTER INSULATION COSMETIC DEPRESSION, LEAD STRETCHED, APPARENT EXPLANT DAMAGE. PROXIMAL SEGMENT RETURNED AND ANALYZED. (B)(4): NO ANOMALIES FOUND, OUTER INSULATION COSMETIC ESC. PROXIMAL SEGMENT RETURNED AND ANALYZED. (B)(4): NO ANOMALIES FOUND, INNER INSULATION PULLED APART (OVERSTRESS), INNER TUBING TORN, OUTER INSULATION COSMETIC DEPRESSION, APPARENT EXPLANT DAMAGE. PROXIMAL SEGMENT RETURNED AND ANALYZED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED WHEN THE NURSE WAS UNABLE TO GET HOLD OF THE PATIENT, THE POLICE CHECKED AND FOUND THE PATIENT DECEASED AT HOME. THE NURSE WAS QUESTIONING WHAT WOULD CAUSE THE IMPEDANCE TO RISE ON EVERY LEAD. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED.
IT WAS REPORTED THE PATIENT WAS FOUND BY POLICE DECEASED AT HOME. THE NURSE WAS QUESTIONING WHAT WOULD CAUSE THE IMPEDANCE TO RISE ON EVERY LEAD. FOLLOW UP REVEALED PATIENT HAD LAST BEEN SEEN A MONTH PRIOR TO DEATH WITH A FULL DEVICE INTERROGATION. THERE WERE NO PROBLEMS WITH DEVICE AND IMPEDANCES ON THE LEADS WERE NORMAL. THEY DO NOT HAVE THE OFFICIAL CAUSE OF DEATH, BUT REPORT PATIENT WAS VERY SICK AND ATTRIBUTE THE DEATH TO MIXED ISCHEMIC AND NO-ISCHEMIC CARDIOMYOPATHY. PATIENT HAD EJECTION FRACTION OF 25 - 30% AND RENAL PROBLEMS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONCERTO CRT-D DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | NIK | MEDTRONIC MED REL, INC. | C154DWK | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 87 YR | Death |