FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES

MDR report key: 19309797 · Received May 14, 2024

Report

Report Number
1024879-2024-00424
Event Type
Malfunction
Date Received
May 14, 2024
Date of Event
April 2, 2024
Report Date
May 22, 2024
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
30382903679837
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY : "MATERIAL #: 367983. LOT/BATCH #: 3153025. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR EVALUATION. THEREFORE, TWENTY (20) RETENTION SAMPLES WERE SUBJECTED TO A DRAW TEST FOR LOW OR NO DRAW AND THE ISSUE RELATING TO UNDERFILLED WAS OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE OF UNDERFILLED. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES, ONE OF THE COLLECTION TUBES HAD INSUFFICIENT NEGATIVE PRESSURE, AND NO PATIENT IMPACT REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE USING BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES, ONE OF THE COLLECTION TUBES HAD INSUFFICIENT NEGATIVE PRESSURE, AND NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
707428 BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO., (BD) 3153025 30382903679837

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown