BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES
Report
- Report Number
- 1024879-2024-00423
- Event Type
- Malfunction
- Date Received
- May 14, 2024
- Date of Event
- April 8, 2024
- Report Date
- May 22, 2024
- Manufacturer
- BECTON, DICKINSON & CO., (BD)
- Product Code
- JKA
- UDI-DI
- 30382903679837
- PMA / PMN Number
- BK050036
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
H.6 INVESTIGATION SUMMARY: MATERIAL #: 367983. LOT/BATCH #: 3153025. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR EVALUATION. THEREFORE, TWENTY (20) RETENTION SAMPLES WERE SUBJECTED TO A DRAW TEST FOR LOW OR NO DRAW AND THE ISSUE RELATING TO UNDERFILLED WAS OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE OF UNDERFILLED. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT WHILE USING BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES, TWO OF THE COLLECTION TUBES HAD INSUFFICIENT NEGATIVE PRESSURE; ONLY ABOUT 1.5 ML OF BLOOD IN ONE TUBE, AND 1 ML IN THE OTHER TUBE. NO PATIENT IMPACT REPORTED.
IT WAS REPORTED THAT WHILE USING BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES, TWO OF THE COLLECTION TUBES HAD INSUFFICIENT NEGATIVE PRESSURE; ONLY ABOUT 1.5 ML OF BLOOD IN ONE TUBE, AND 1 ML IN THE OTHER TUBE. NO PATIENT IMPACT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 708410 | BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON & CO., (BD) | 3153025 | 30382903679837 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |