COLLEAGUE PUMP CE ENGLISH VERSION
Report
- Report Number
- 6000001-2010-05881
- Event Type
- Malfunction
- Date Received
- December 16, 2010
- Date of Event
- November 15, 2010
- Report Date
- November 22, 2010
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE CONDITION OF UPSTREAM OCCLUSION ALARM WAS CONFIRMED DUE TO A FAULTY PUMP HEAD MODULE. THE PUMP HEAD MODULE WAS REPLACED TO CORRECT THE REPORTED CONDITION. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. A SERVICE REVIEW WAS PERFORMED AND THERE ARE NO SERVICE ACTIVITIES IN THE PREVIOUS 12 MONTHS FOR THIS DEVICE. THIS DEVICE UTILIZES USER INTERFACE MODULE MASTER SOFTWARE VERSION 5.09.90 CATEGORIZED AS REMEDIATED. (B)(4)
A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT HAVE A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.(B)(4)
THIS IS A REPORT FROM BAXTER (B)(4) OF AN UPSTREAM OCCLUSION ALARM EVERY 20 MINUTES ON A COLLEAGUE INFUSION PUMP. IT IS UNKNOWN WHEN THE REPORTED CONDITION OCCURRED. IT IS UNKNOWN IF PATIENT INJURY OR MEDICAL INTERVENTION OCCURRED. THE SOFTWARE VERSION FOR THIS DEVICE IS CURRENTLY NOT KNOWN.