FDA Adverse Event Injury Summary report: N

LOGIC CR TIB INSERT STD, SZ 3, 15MM

MDR report key: 19308824 · Received May 14, 2024

Report

Report Number
1038671-2024-01226
Event Type
Injury
Date Received
May 14, 2024
Date of Event
September 22, 2023
Report Date
November 14, 2024
Manufacturer
EXACTECH, INC.
Product Code
JWH
UDI-DI
10885862174376
PMA / PMN Number
K111400
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

D10: CONCOMITANT DEVICES: 4041357 521-78-31 - THREADED PIN SIZE 2.6 COLLARLESS 2PK. 4329435 521-78-23 - THREADED PIN SIZE 2.3 COLLARED 2PK. 4526879 02-010-03-0330 - LOGIC CR FEMORAL CEM, RIGHT, SZ 3. 4556114 200-02-35 - THREE PEG PATELLA 35MM. 4631894 02-012-45-3030 - LGC TIBIAL FIT TRAY CEM SZ 3F / 3T. THE DEVICE WAS NOT RETURNED FOR EVALUATION AND NO PHOTOGRAPHS OR X-RAYS WERE PROVIDED FOR REVIEW; THEREFORE THE REPORTED EVENT CANNOT BE CONFIRMED THROUGH ANALYSIS. NO OPERATIVE NOTES WERE PROVIDED; THEREFORE, A REVIEW OF DEVICE USAGE AND TECHNIQUE COULD NOT BE PERFORMED. NO INFORMATION CONCERNING PATIENT CONDITIONS, CO-MORBIDITIES, OR OTHER HEALTH OR ANATOMICAL CONCERNS WAS PROVIDED AND IT IS THEREFORE UNKNOWN IF PATIENT-RELATED FACTORS MAY HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE CAUSE OF THE REPORTED EVENT CANNOT BE DETERMINED BASED ON THE INFORMATION MADE AVAILABLE. SHOULD ADDITIONAL, RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Additional Manufacturer Narrative · 0

H6: CORRECTED THE FOLLOWING: HEALTH EFFECT CLINICAL CODE, TYPE OF INVESTIGATION THE REASON FOR THE REVISION REPORTED CANNOT BE CONFIRMED FROM THE INFORMATION PROVIDED BUT MAY BE THE RESULT OF PROSTHESIS WEAR OR DUE TO INCLUSION OF THE POLYETHYLENE IN THE PACKAGING RECALL. ADDITIONALLY, A CONTRIBUTING FACTOR TO THE REPORTED WEAR MAY HAVE BEEN THE RESULT OF BEING PACKAGED IN A NON-CONFORMING VACUUM BAG FOR MORE THAN FIVE YEARS. POTENTIAL CONTRIBUTIONS OF USER AND PATIENT-RELATED CONSIDERATIONS TO THE EVENT COULD NOT BE ASSESSED AS THE DEVICES WERE NOT AVAILABLE FOR EVALUATION AND IMAGES, RADIOGRAPHS, AND RELEVANT CLINICAL INFORMATION WERE NOT PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 82 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS POLYETHYLENE WEAR, PAIN, SWELLING, AND INSTABILITY. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
767722 LOGIC CR TIB INSERT STD, SZ 3, 15MM PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL JWH EXACTECH, INC. 10885862174376

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention SEE H10.