STRATA NSC VALVE, REGULAR
Report
- Report Number
- 2021898-2010-00350
- Event Type
- Injury
- Date Received
- December 14, 2010
- Date of Event
- November 9, 2010
- Report Date
- November 16, 2010
- Manufacturer
- MEDTRONIC NEUROSURGERY
- Product Code
- JXG
- PMA / PMN Number
- K033850
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE DEVICE WAS PATENT, BUT IT DID NOT MEET THE REQUIREMENTS FOR LEAK TESTING DUE TO MULTIPLE TEARS IN THE SILICONE DOME ABOVE THE RESERVOIR. IT IS UNK HOW OR WHEN THESE DAMAGES OCCURRED. THERE WERE ALSO SCRATCHES ON THE BASE OF THE RESERVOIR, BELOW THE TEARS. THEREFORE, IT IS POSSIBLE THAT THE DAMAGES OCCURRED WHEN INJECTING OR CSF SAMPLING FROM THE VALVE. THE INSTRUCTIONS FOR USE THAT ACCOMPANY THE DEVICE NOTE THAT INJECTION OR CSF SAMPLING THROUGH THE DOME IS ALLOWED BY USE OF A 25-GAUGE OR SMALLER NON-CORING NEEDLE. IT IS ALSO CAUTIONED THAT "LOW TEAR STRENGTH IS A CHARACTERISTIC OF UNREINFORCED SILICONE ELASTOMER MATERIALS. CARE MUST BE TAKEN ON INSERTION AND REMOVAL OF THE NEEDLE." THE VALVE DID NOT MEET THE REQUIREMENTS FOR PRESSURE-FLOW AND PREIMPLANTATION TESTING. PROTEINACEOUS DEBRIS WAS NOTED IN THE VALVE MECHANISM. DEBRIS WITHIN THE VALVE MAY HOLD PRESSURE-FLOW CONTROLLING MECHANISMS OPEN, RESULTING IN OVERDRAINAGE AND REFLUX. A REVIEW OF THE MFG RECORDS SHOWED NO ANOMALIES. NO PT IMPACT WAS REPORTED. ALL OF OUR VALVES ARE 100% TESTED AT THE TIME OF MFR.
IT WAS REPORTED THAT A FEW DAYS AFTER THE SURGERY, THE PT'S SYMPTOMS HAD NOT IMPROVED, AND SUBCUTANEOUS FLUID WAS FOUND BY A CT SCAN. ON (B)(6) 2010, THE PT HAD A REVISION SURGERY AND THE DOCTOR FOUND THE RESERVOIR HAD A LEAK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STRATA NSC VALVE, REGULAR | JXG | MEDTRONIC NEUROSURGERY | NA | C53173 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization| R | CATALOG# 91503, LOT# C60995| CATALOG# 93522, LOT# C64800 |