FDA Adverse Event Injury Summary report: N

DELTA SHUNT KIT REGULAR, PL 1.5

MDR report key: 1930698 · Received December 14, 2010

Report

Report Number
2021898-2010-00348
Event Type
Injury
Date Received
December 14, 2010
Date of Event
September 20, 2010
Report Date
November 16, 2010
Manufacturer
MEDTRONIC NEUROSURGERY
Product Code
JXG
PMA / PMN Number
K902783
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS PATENT AND PASSED SIPHON AND REFLUX TESTING. IT DID NOT MEET THE REQUIREMENTS FOR LEAK TESTING DUE TO MULTIPLE TEARS IN THE SILICONE DOME ABOVE THE RESERVOIR. IT IS UNK HOW OR WHEN THESE DAMAGES OCCURRED. THERE WERE ALSO SCRATCHES ON THE BASE OF THE RESERVOIR, BELOW THE TEARS. THEREFORE, IT IS POSSIBLE THAT THE DAMAGES OCCURRED WHEN INJECTING OR CSF SAMPLING FROM THE VALVE. THE INSTRUCTIONS FOR USE THAT ACCOMPANY THE DEVICE NOTE THAT INJECTION OR CSF SAMPLING THROUGH THE DOME IS ALLOWED BY USE OF A 25-GAUGE OR SMALLER NON-CORING NEEDLE. IT IS ALSO CAUTIONED THAT "LOW TEAR STRENGTH IS A CHARACTERISTIC OF UNREINFORCED SILICONE ELASTOMER MATERIALS. CARE MUST BE TAKEN ON INSERTION AND REMOVAL OF THE NEEDLE." THE DEVICE PERFORMANCE DID NOT MEET ESTABLISHED SPECIFICATIONS FOR PRESSURE-FLOW AND PREIMPLANTATION TESTING. PROTEINACEOUS DEBRIS WAS NOTED ON THE VALVE MEMBRANE. DEBRIS WITHIN THE VALVE MAY HOLD PRESSURE-FLOW CONTROLLING MECHANISMS OPEN, RESULTING IN OVERDRAINAGE. A REVIEW OF THE MFG RECORDS SHOWED NO ANOMALIES. NO PT IMPACT WAS REPORTED. ALL OF OUR VALVES ARE 100% TESTED AT THE TIME OF MFR.

Description of Event or Problem · 1

IT WAS REPORTED THAT ONE WEEK AFTER THE SURGERY, THE PT FELT UNCOMFORTABLE. A CT REVIEW FOUND A SMALL AMOUNT OF FLUID AROUND THE VALVE, AND THE VENTRICLES WERE ENLARGED. THE DOCTOR PERFORMED A SECOND SURGERY AND FOUND THAT THE RESERVOIR WAS LEAKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DELTA SHUNT KIT REGULAR, PL 1.5 JXG MEDTRONIC NEUROSURGERY NA C66509

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization| R