FDA Adverse Event Malfunction Summary report: N

AUTOMATED PD SET W/CASSETTE4 PRONG

MDR report key: 1930635 · Received December 16, 2010

Report

Report Number
1423500-2010-06952
Event Type
Malfunction
Date Received
December 16, 2010
Date of Event
November 21, 2010
Report Date
November 21, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT IS FOR A REPORT OF A PATIENT DISCONNECTING AND NOT CAPPING OFF THE PATIENT LINE. THIS COMPLAINT CANNOT BE CONFIRMED IN THE LAB DUE TO A LACK OF SAMPLE. THE NOT ENOUGH DATA ARE AVAILABLE WITHIN THE COMPLAINT INFORMATION TO IDENTIFY ROOT CAUSE; THEREFORE THE ROOT CAUSE OF THE COMPLAINT WAS NOT DETERMINED. THIS REVIEW FOUND THE LABELING ADEQUATE FOR THE USE ERROR IN THE COMPLAINT. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS DISCARDED. SHOULD ANY ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. THE LOT NUMBER WAS NOT PROVIDED; THEREFORE, A BATCH REVIEW CANNOT BE CONDUCTED.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A NEED TO END THERAPY, WHICH OCCURRED ON THE HOMECHOICE (HC) DURING USE DURING DWELL 1. THE HOME PATIENT (HP) STATED THAT THEY HAD TO DISCONNECT FROM THE MACHINE BECAUSE OF AN EMERGENCY AND NOW THE HP WAS READY TO RECONNECT. THE LINE WAS NOT CAPPED. THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) ADVISED THE HP THAT THEY WOULD NEED TO END THERAPY AND START OVER WITH NEW SUPPLIES. THE TSR ASSISTED IN ENDING THERAPY. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. PRODUCT SURVEILLANCE CONTACTED THE HOME PATIENT (HP) ON (B)(4) 2010 REGARDING THE REPORTED PROBLEM. THE HP HAS CONTINUED THERAPY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED AS A RESULT OF THIS INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATED PD SET W/CASSETTE4 PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 33 YR PD CYCLER