FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1930609
·
Received December 1, 2010
Report
- Report Number
- 2531779-2010-02630
- Event Type
- Malfunction
- Date Received
- December 1, 2010
- Date of Event
- November 1, 2010
- Report Date
- November 1, 2010
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Removal / Correction Number
- 2531779-03/24/2010-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED. THE LOSS OF DETECTION WAS DUPLICATED DURING TESTING. INVESTIGATION REVEALED A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. THE FORCE SENSOR FLEX PIN WAS FOUND TO BE BROKEN OFF IN THE FORCE SENSOR SOCKET.
Description of Event or Problem · 1
AN INTERMITTENT FORCE SENSOR CONNECTION AND A BROKEN FORCE SENSOR PIN WAS DISCOVERED DURING INVESTIGATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 109 YR |