FDA Adverse Event Malfunction Summary report: N

AUTOMATED PD SET W/CASSETTE4 PRONG

MDR report key: 1930589 · Received December 16, 2010

Report

Report Number
1423500-2010-06944
Event Type
Malfunction
Date Received
December 16, 2010
Date of Event
November 21, 2010
Report Date
November 24, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING INVESTIGATION BY BAXTER THIS INCIDENT WAS DETERMINED TO BE CAUSED BY USE/USER ERROR AND THERE WAS NO ALLEGATION OF A PRODUCT MALFUNCTION. THEREFORE, THE DEVICE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF BAXTER'S INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR 2240 (AIR IN SET) WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. AS PER THE INVESTIGATION, CAUSE OF THE SE 2240 IS DUE TO AIR BEING SUCKED INTO THE DISPOSABLE AFTER THE SUPPLY BAG FELL AND DISCONNECTED. A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED WITH EXCEPTION ISSUES NOTED DURING THE MANUFACTURING PROCESS FOR EXTRA SHEETING ON CASSETTE. QUALITY RE-INSPECTION WAS ACCEPTABLE. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

A PATIENT'S RELATIVE CONTACTED BAXTER REGARDING ASSISTANCE WITH A SYSTEM ERROR 2240 WHILE USING THE HOMECHOICE (HC) DURING AN UNSPECIFIED CYCLE. THE SUPPLY BAG HAD FALLEN DOWN BY MISTAKE (DUE TO USER ERROR). THE PATIENT ELECTED TO FINISH THERAPY MANUALLY. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATED PD SET W/CASSETTE4 PRONG SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME H10E24056

Patients

Seq Age Sex Outcome Treatment
1