FDA Adverse Event Malfunction Summary report: N

AC POWER MODULE

MDR report key: 1930508 · Received November 30, 2010

Report

Report Number
1218950-2010-02411
Event Type
Malfunction
Date Received
November 30, 2010
Report Date
November 4, 2010
Manufacturer
PHILIPS HEALTHCARE
Product Code
MKJ
PMA / PMN Number
K031187
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): IT WAS REPORTED TO PHILIPS THAT THE AC POWER MODULE FAILED. THE UNIT AND AC POWER MODULE WERE EVALUATED BY A PHILIPS FIELD SERVICE ENGINEER AND THE FAILURE WAS VERIFIED. REPLACEMENT OF THE MODULE RESOLVED THE FAILURE. THE UNIT PASSED ALL POST-SERVICING TESTING. THE UNIT REMAINS AT THE CUSTOMER SITE WITH THE NEW MODULE INSTALLED.

Description of Event or Problem · 1

IT WAS REPORTED TO PHILIPS THAT THE AC POWER MODULE FAILED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AC POWER MODULE MKJ PHILIPS HEALTHCARE M3539A

Patients

Seq Age Sex Outcome Treatment
1