FDA Adverse Event
Malfunction
Summary report: N
R/R SINGLE TRIG PIN COLLET
MDR report key: 1930483
·
Received November 30, 2010
Report
- Report Number
- 1811755-2010-01842
- Event Type
- Malfunction
- Date Received
- November 30, 2010
- Date of Event
- November 9, 2010
- Report Date
- November 9, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE PIN.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LOANER DEVICE WOULD NOT HOLD SMALL GAGE PIN. THE CONDITION OF THE DEVICE WAS DISCOVERED DURING ROUTINE MAINTENANCE IN HOUSE. THERE WAS NO PT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | R/R SINGLE TRIG PIN COLLET | INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR | KIJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |