FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1930479 · Received November 30, 2010

Report

Report Number
2531779-2010-02639
Event Type
Malfunction
Date Received
November 30, 2010
Report Date
November 2, 2010
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL .THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE KEYPAD APPEARED TO BE INTACT WITH NO LIFTING OR PEELING OBSERVED, THE "UP/CONTRAST" KEYPAD BUTTONS DID NOT RESPOND TO USER INPUTS, THE "DOWN/OK" KEYPAD BUTTONS INTERMITTENTLY RESPONDED TO USER INPUTS, IT WAS OBSERVED THAT THE "UP/CONTRAST" KEY CONTACTS TRACES WERE BROKEN, THE "DOWN/OK" KEY CONTACTS BECAME INVERTED WHEN PRESSED AND THERE WAS EVIDENCE OF ADHESIVE/CONTAMINATION UNDER ALL THE KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE KEYPAD BUTTONS REPORTEDLY NOT RESPONDING WHEN PRESSED. IT WAS REPORTED THAT THEY ALL FEEL AND SPRING BACK NORMALLY WHEN PRESSED AND THAT THE KEYPAD IS INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING NA