6 FR T/L POWERPICC SOLO NURSE BASIC WITH MI AND TLS
Report
- Report Number
- 3006260740-2010-00343
- Event Type
- Malfunction
- Date Received
- November 18, 2010
- Date of Event
- October 28, 2010
- Report Date
- October 29, 2010
- Manufacturer
- C. R. BARD INC. (BASD)
- Product Code
- LJS
- PMA / PMN Number
- K072230
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- NURSE
Narratives
THE COMPLAINT THAT THE DISTAL TIP OF THE TLS STYLISTS DETACHED WAS CONFIRMED. THE PRODUCTS RETURNED FOR EVAL WERE TWO TLS STYLETS WITH A BREAK WITHIN THE DISTAL MAGNET REGION AND TWO SMALL SEGMENTS OF THE TLS STYLETS. THE PROXIMAL SEGMENTS INCLUDED THE YELLOW PULL-TAB, THE T-LOCK ASSEMBLY, THE CENTRAL CORE WIRE AND SOME OF THE MAGNETIC REGION. MULTIPLE KINKS WERE SEEN IN THE MAGNETIC REGIONS OF THE STYLETS. ONE OF THE SAMPLES CONTAINED A SMALL SECTION OF STRETCHED EMPTY POLYIMIDE TUBING. THE BREAK ON THIS SAMPLE WAS INTRAMAGNETIC, BREAKING THE DISTAL MAGNET. THE MAGNET WAS SLIGHTLY RECESSED WITHIN THE TUBING AND THE POLYIMIDE TUBING WAS BENT OVER THE MAGNET. THE BREAK ON THE OTHER SAMPLE WAS AN INTERMAGNETIC BREAK WITH THE DISTAL END OF THE MAGNET PROTRUDING SLIGHTLY OUT OF THE POLYIMIDE TUBING. THE SMALLER RETURNED SEGMENTS WERE FROM THE MAGNETIC REGION OF THE TLS STYLETS. ONE OF THE SMALL SEGMENTS WAS BROKEN ON BOTH ENDS. THE SEGMENT CONTAINED SEVEN COMPLETE MAGNETS AND ONE PARTIAL MAGNET. THIS SEGMENT WAS 0.72" LONG. THE MAGNETS WERE RECESSED WITHIN THE POLYIMIDE TUBING ON ONE SIDE AND PROTRUDING FROM THE TUBING ON THE OTHER SIDE. THE TUBING WAS BENT OVER THE RECESSED MAGNET. THE OTHER MAGNETIC REGION PIECE CONTAINED 6 MAGNETS AND THE DISTAL EPOXY TIP. THIS SEGMENT WAS 0.68" LONG. THE PROXIMAL MAGNET WAS SLIGHTLY RECESSED WITHIN THE TUBING AND THE BLUE SHRINK TUBING WAS BENT OVER THE END OF THE MAGNET. THE POLYIMIDE TUBING WAS BENT OUT OF ROUND. THE PRODUCT IFU WARNS, "ENSURE THAT THE STYLET TIP DOES NOT EXTEND BEYOND THE TRIMMED END OF THE CATHETER. EXTENSION OF THE STYLET TIP BEYOND THE CATHETER END, COMBINED WITH KINKING AND EXCESSIVE FORCES MAY RESULT IN VESSEL DAMAGE, STYLET DAMAGE, DIFFICULT REMOVAL, STYLET TIP SEPARATION, POTENTIAL EMBOLISM AND RISK OF PT INJURY." A CHR OF LOT # REUC0757 SHOWED NO OTHER SIMILAR PRODUCT COMPLAINTS FROM THIS LOT NUMBER.
IT WAS REPORTED THAT THE DOCTOR ADVANCED THE CATHETER AND IT WENT INTO THE SIDE VESSEL. CONTINUED TO ADVANCE THINKING SHE WAS IN BASILIC AND KINKED THE CATHETER. PULLED CATHETER BACK AND OUT. NURSE PULLED STYLET BACK AND FLUSHED, BUT NOTICED STYLET TIP VISIBLE AT END. PULLED THE STYLET WIRE FROM DISTAL END AND FOUND THE TLS STYLET HAD KINKED AND BROKEN INTO TWO PIECES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 6 FR T/L POWERPICC SOLO NURSE BASIC WITH MI AND TLS | LJS | C. R. BARD INC. (BASD) | REUC0757 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |