UNKNOWN ZIMMER IMPACTOR HANDLE
Report
- Report Number
- 1822565-2010-00837
- Event Type
- Malfunction
- Date Received
- October 5, 2010
- Date of Event
- June 22, 2010
- Report Date
- August 26, 2010
- Manufacturer
- ZIMMER, INC.
- Product Code
- LXH
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4): EVALUATION SUMMARY: THE ITEM NUMBER AND LOT NUMBER OF THE IMPACTOR HANDLE USED WITH THE POSITIONER HEAD IS UNKNOWN. THE TENTATIVE FIELD AGE OF THE POSITIONED HEAD IS 22 YRS. ONE OR A COMBINATION OF THE FOLLOWING FACTORS MIGHT HAVE CONTRIBUTED TO THE FAILURE: INSUFFICIENT THREAD ENGAGEMENT DURING IMPACTION MAY HAVE LED TO HIGH STRESS ON THE THREADS AND SUBSEQUENT FAILURE; OFF-AXIS IMPACTION OF THE COMPONENT EITHER WITH SUFFICIENT OR INSUFFICIENT THREAD ENGAGEMENT; THE DEVICE WAS INADVERTENTLY DROPPED OR DAMAGED DURING CLEANING/AUTOCLAVE. HOWEVER, NO DEFINITIVE CAUSE ANALYSIS CAN BE PERFORMED WITH CERTAINTY. EVALUATION: REVIEW OF THE DEVICE HISTORY RECORDS WAS NOT POSSIBLE AS THE PRODUCT AND LOT NUMBER REQUIRED FOR RETRIEVAL WERE UNAVAILABLE. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE AVAILABLE INFORMATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL SUBSTANTIVE INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER, INC CONSIDERS THE INVESTIGATION CLOSED.
IT IS REPORTED THAT DURING LEFT TOTAL HIP ARTHROPLASTY BEING PERFORMED ON A PATIENT WITH AVASCULAR NECROSIS, THE HEAD OF A ZIMMER IMPACTOR BROKE OFF FROM HANDLE. THE SCREW HOLDING THE FLAT ROUND HEAD ON THE HANDLE APPEARED TO BE BROKEN. X-RAYS TAKEN AND NO METAL FRAGMENTS WERE NOTED TO BE RETAINED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNKNOWN ZIMMER IMPACTOR HANDLE | HIP INSTRUMENT | LXH | ZIMMER, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 24 YR |