FDA Adverse Event Malfunction Summary report: N

PROXIMATE** RELOADABLE LINEAR CUTTER WITH SAFETY LOCK-OUT

MDR report key: 1930105 · Received December 16, 2010

Report

Report Number
3005075853-2010-07189
Event Type
Malfunction
Date Received
December 16, 2010
Date of Event
November 30, 2010
Report Date
December 1, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH NO RELOAD PRESENT. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED AND CUT WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. EVENT COULD NOT BE CONFIRMED AS THE FIRING KNOB WORKED WITHOUT ANY DIFFICULTIES NOTED DURING THE FUNCTIONAL TEST. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4).INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ADDITIONAL INFORMATION: THERE WERE NO DIFFICULTY ON THE FIRING. THE DEVICE WAS NOT FIRED ACROSS THE HARD OBJECT. THE TARGET TISSUE WAS CLAMPED EVENLY IN THE JAW AND WAS NOT SO THICK.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A GASTRECTOMY PROCEDURE, THE FIRING KNOB WAS HARD AND DIFFICULT TO RETURN TO THE HOME POSITION AFTER THE FIRST FIRING. THE KNOB WAS RETURNED TO THE HOME POSITION FORCIBLY. SOME STAPLES WERE UNFORMED. REINFORCEMENT MATERIAL WAS NOT USED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROXIMATE** RELOADABLE LINEAR CUTTER WITH SAFETY LOCK-OUT STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA G4TL2H

Patients

Seq Age Sex Outcome Treatment
1 CARTRIDGE