FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 1930086 · Received December 2, 2010

Report

Report Number
1720753-2010-04388
Event Type
Malfunction
Date Received
December 2, 2010
Date of Event
November 5, 2010
Report Date
December 1, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE CONTACTS ON THE POWER DISTRIBUTION BOARD WERE REPAIRED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM FAILED TO BOOT UP. NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9600 FLUOROSCOPIC XRAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9600

Patients

Seq Age Sex Outcome Treatment
1