FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 1930039 · Received December 2, 2010

Report

Report Number
1720753-2010-04379
Event Type
Malfunction
Date Received
December 2, 2010
Date of Event
November 15, 2010
Report Date
November 30, 2010
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND RESEATED THE 5 VOLT POWER SUPPLY CONNECTORS AND ADJUSTED THE 5 VOLT POWER SUPPLY. SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

CUSTOMER REPORTED THE SYSTEM DISPLAYED A COMMUNICATION ERROR. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1