FDA Adverse Event Malfunction Summary report: N

IP PTZ ERGOJUST VIDEO EXTENSIO

MDR report key: 19296674 · Received May 13, 2024

Report

Report Number
9612330-2024-00025
Event Type
Malfunction
Date Received
May 13, 2024
Date of Event
February 6, 2024
Report Date
August 22, 2024
Manufacturer
NATUS MEDICAL INCORPORATED
Product Code
OMB
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

FOLLOW UP REPORT 002 REF NATUS COMPLAINT# (B)(4). IT WAS CONFIRMED THE CUSTOMER WILL NOT BE RETURNING THE CART FOR EVALUATION. ENGINEERING REVIEW STATES,"IT'S DIFFICULT TO PUSH THE CART AROUND A CORNER WHEN IT DOESN'T HAVE TWO DIRECTIONAL LOCKING CASTERS AT THE REAR. ECR#22803 WAS RELEASED FOR THE IMPLEMENTATION OF TWO DIRECTIONAL LOCKING CASTERS FOR EASIER MOBILITY AROUND CORNERS. (RELATED ECO#44325 ALSO GENERATED). PER (B)(4) REV 70 XLTEK EEG/PSG RISK ANALYSIS SPREADSHEET, HAZARD ID 6.6 CAUSE - ERGONOMIC DESIGN OF SYSTEM RESULTS IN REPETITIVE STRESS INJURY. EFFECTS (HARM) -RANGE FROM CLINICALLY INSIGNIFICANT TO IRREVERSIBLE INJURY RESULTING FROM DIRECT PHYSICAL INJURY RISK / BENEFIT ANALYSIS - LOW THE HAZARDS IDENTIFIED HAVE BEEN REDUCED AS FAR AS POSSIBLE, AND THE RESIDUAL RISK FOR THESE HAZARDS ARE CATEGORIZED AS HAVING AN RBA RATING OF LOW DUE TO THE NATURE OF THE HAZARD. HAZARDS HAVE ASSOCIATED WARNINGS DISCLOSED IN THE IFU. RISK OUTWEIGHED BY BENEFIT OF USE OF DEVICE. NO RELATED CAPAS. FURTHER REVIEW AND APPROVAL OF INVESTIGATION DETAILS TO BE COMPLETED BEFORE FINAL CLOSURE OF THIS CASE.

Additional Manufacturer Narrative · 0

INITIAL REPORT REF NATUS COMPLAINT#(B)(4). IT WAS STATED BY THE CUSTOMER THAT THE ISSUE IS NOT LINKED TO A SPECIFIC CART. DEVICE SERIAL /LOT NUMBERS PROVIDED WERE: (B)(6) PHOTOGRAPH OF A DEVICE WAS PROVIDED BY THE CUSTOMER, SHOWING BOX OBSTRUCTING COLLEAGUE VIEW. FURTHER INVESTIGATION TO BE CARRIED OUT.

Additional Manufacturer Narrative · 0

FOLLOW UP REPORT 001 REF NATUS COMPLAINT (B)(4) INSTALL DATE 17-AUG-2020 FURTHER INVESTIGATION TO BE CARRIED OUT.

Additional Manufacturer Narrative · 0

NATUS COMPLAINT#: (B)(4) WAS CLOSED AUG 21, 2024. FAILURE CONFIRMED: NO. INVESTIGATION RESULT CODE: NEURO SBU/PRODUCT NOT RETURNED. ECR#: 22803 WAS RELEASED FOR THE IMPLEMENTATION OF TWO DIRECTIONAL LOCKING CASTERS FOR EASIER MOBILITY AROUND CORNERS. (RELATED ECO#: 44325 WAS ALSO GENERATED).

Description of Event or Problem · 0

IP PTZ ERGOJUST VIDEO EXTENSION - CUSTOMER STATED WITH REPETITIVE PUSHING OF CART WITH VIEW OBSTRUCTED BY MOUNTED ELECTRONIC BOX APPROXIMATELY 10.5 X 11.25 INCHES, COLLEAGUE IS FORCED TO MAINTAIN A NON-ERGONOMIC (AWKWARD & BENT/TWISTED) POSTURE WHEN PUSHING CART TO SEE WHAT IS IN FRONT AND PREVENT FROM BUMPING INTO ANYTHING WHEN MOVING CART FROM PLACE TO PLACE. ERGONOMIC EVALUATION COMPLETED BY THE CUSTOMER ON 02/12/2024, SUPPORTING THAT COLLEAGUE INJURY IS RELATED TO CART LAYOUT. RESTRICTIONS AND PHYSICAL THERAPY WAS RECOMMENDED. 4 CARTS IDENTIFIED. DOCUMENTING CART 1/4.

Description of Event or Problem · 0

IP PTZ ERGOJUST VIDEO EXTENSION - CUSTOMER STATED WITH REPETITIVE PUSHING OF CART WITH VIEW OBSTRUCTED BY MOUNTED ELECTRONIC BOX APPROXIMATELY 10.5 X 11.25 INCHES, COLLEAGUE IS FORCED TO MAINTAIN A NON-ERGONOMIC (AWKWARD & BENT/TWISTED) POSTURE WHEN PUSHING CART TO SEE WHAT IS IN FRONT AND PREVENT FROM BUMPING INTO ANYTHING WHEN MOVING CART FROM PLACE TO PLACE. ERGONOMIC EVALUATION COMPLETED BY THE CUSTOMER ON 02/12/2024, SUPPORTING THAT COLLEAGUE INJURY IS RELATED TO CART LAYOUT. RESTRICTIONS AND PHYSICAL THERAPY WAS RECOMMENDED.

Description of Event or Problem · 0

IP PTZ ERGOJUST VIDEO EXTENSION - CUSTOMER STATED WITH REPETITIVE PUSHING OF CART WITH VIEW OBSTRUCTED BY MOUNTED ELECTRONIC BOX APPROXIMATELY 10.5 X 11.25 INCHES, COLLEAGUE IS FORCED TO MAINTAIN A NON-ERGONOMIC (AWKWARD & BENT/TWISTED) POSTURE WHEN PUSHING CART TO SEE WHAT IS IN FRONT AND PREVENT FROM BUMPING INTO ANYTHING WHEN MOVING CART FROM PLACE TO PLACE. ERGONOMIC EVALUATION COMPLETED BY THE CUSTOMER ON 02/12/2024, SUPPORTING THAT COLLEAGUE INJURY IS RELATED TO CART LAYOUT. RESTRICTIONS AND PHYSICAL THERAPY WAS RECOMMENDED. 4 CARTS IDENTIFIED. DOCUMENTING CART 1/4

Description of Event or Problem · 0

IP PTZ ERGOJUST VIDEO EXTENSION: CUSTOMER STATED WITH REPETITIVE PUSHING OF CART WITH VIEW OBSTRUCTED BY MOUNTED ELECTRONIC BOX APPROXIMATELY 10.5 X 11.25 INCHES, COLLEAGUE IS FORCED TO MAINTAIN A NON-ERGONOMIC (AWKWARD & BENT/TWISTED) POSTURE WHEN PUSHING CART TO SEE WHAT IS IN FRONT AND PREVENT FROM BUMPING INTO ANYTHING WHEN MOVING CART FROM PLACE TO PLACE. ERGONOMIC EVALUATION COMPLETED BY THE CUSTOMER ON (B)(6) 2024, SUPPORTING THAT COLLEAGUE INJURY IS RELATED TO CART LAYOUT. RESTRICTIONS AND PHYSICAL THERAPY WAS RECOMMENDED. 4 CARTS IDENTIFIED. DOCUMENTING CART 1/4.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
411441 IP PTZ ERGOJUST VIDEO EXTENSIO IP PTZ ERGOJUST VIDEO EXTENSIO OMB NATUS MEDICAL INCORPORATED IP-EC-VIDEO-EXT

Patients

Seq Age Sex Outcome Treatment
1 63 YR Female Other