FDA Adverse Event Injury Summary report: N

ENDOWRIST

MDR report key: 19292846 · Received May 10, 2024

Report

Report Number
2955842-2024-14288
Event Type
Injury
Date Received
May 10, 2024
Date of Event
April 12, 2024
Report Date
April 16, 2024
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
10886874114817
PMA / PMN Number
K162973
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE THE SUCTION IRRIGATOR INSTRUMENT TO PERFORM FAILURE ANALYSIS. THE INSTRUMENT WAS ANALYZED AND FOUND TO HAVE A PIECE MEASURING APPROXIMATELY 0.089" X 0.203" BROKEN OFF AT THE DISTAL END. COMPONENTS ADJACENT TO THIS BROKEN GRIP DO NOT SHOW DAMAGE. THE BROKEN PIECE WAS NOT RETURNED WITH THE INSTRUMENT.

Additional Manufacturer Narrative · 0

A RETURN MATERIAL AUTHORIZATION (RMA) HAD BEEN ISSUED REQUESTING TO HAVE THE INTUITIVE SURGICAL, INC. (ISI) DEVICE RETURNED; HOWEVER, ISI HAS NOT RECEIVED THE RMA TO CONFIRM/IDENTIFY THE FAILURE MODE. ADDITIONAL INFORMATION IS BEING GATHERED TO DETERMINE THE CONTRIBUTION OF THE DEVICE TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE PROVIDED IMAGES WAS PERFORMED BY AN ISI FAILURE ANALYSIS ENGINEER, AND THE LOCATION OF WHERE THE FRAGMENT CAME FROM COULD NOT BE DETERMINED BASED ON THE PHOTOS PROVIDED.

Description of Event or Problem · 0

REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED LOW ANTERIOR RESECTION SURGICAL PROCEDURE, THE CUSTOMER USED A SUCTION IRRIGATOR INSTRUMENT, AND A TIP OF PLASTIC BROKE OFF. THE INSTRUMENT WAS STILL FUNCTIONAL, AND THE BROKEN PIECE WAS RETRIEVED AND THROWN IN THE GARBAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2530507 ENDOWRIST SUCTION IRRIGATOR NAY INTUITIVE SURGICAL, INC 480299-06 K16230727 0144 10886874114817

Patients

Seq Age Sex Outcome Treatment
1 NA Prefer Not To Disclose Required Intervention DA VINCI INSTRUMENTS AND ACCESSORIES