FDA Adverse Event Malfunction Summary report: N

I-STAT CTNI CARTRIDGE

MDR report key: 1928451 · Received November 24, 2010

Report

Report Number
2245578-2010-00154
Event Type
Malfunction
Date Received
November 24, 2010
Date of Event
July 19, 2010
Report Date
November 24, 2010
Manufacturer
ABBOTT POINT OF CARE CANADA LTD
Product Code
MMI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). APOC CONDUCTED AND INTERNAL REVIEW OF QUALITY DATA AND CUSTOMER-RETURNED PRODUCT TESTING INDICATE THAT A SECTION OF CTNI LOT S10130A IS ASSOCIATED WITH ELEVATED RATES OF QCC 149. QUALITY ACCEPTANCE CRITERIA IS BEING EVALUATED/UPDATED TO ENSURE APOC INSPECTION CAPTURES ELEVATES QCC RATES.

Description of Event or Problem · 1

ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY THE CUSTOMER STATING THE I-STAT CTNI CARTRIDGES WERE YIELDING AN INCREASE IN THE NUMBER OF QUALITY CHECK CODE (QCC) 149. BASED ON THE INFO PROVIDED BY THE CUSTOMER, THERE WAS NO INDICATION OF PT IMPACT DUE TO INCREASE IN QCC 149S.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 I-STAT CTNI CARTRIDGE CTNI CARTRIDGE MMI ABBOTT POINT OF CARE CANADA LTD S10130A