FDA Adverse Event
Malfunction
Summary report: N
I-STAT CTNI CARTRIDGE
MDR report key: 1928451
·
Received November 24, 2010
Report
- Report Number
- 2245578-2010-00154
- Event Type
- Malfunction
- Date Received
- November 24, 2010
- Date of Event
- July 19, 2010
- Report Date
- November 24, 2010
- Manufacturer
- ABBOTT POINT OF CARE CANADA LTD
- Product Code
- MMI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). APOC CONDUCTED AND INTERNAL REVIEW OF QUALITY DATA AND CUSTOMER-RETURNED PRODUCT TESTING INDICATE THAT A SECTION OF CTNI LOT S10130A IS ASSOCIATED WITH ELEVATED RATES OF QCC 149. QUALITY ACCEPTANCE CRITERIA IS BEING EVALUATED/UPDATED TO ENSURE APOC INSPECTION CAPTURES ELEVATES QCC RATES.
Description of Event or Problem · 1
ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY THE CUSTOMER STATING THE I-STAT CTNI CARTRIDGES WERE YIELDING AN INCREASE IN THE NUMBER OF QUALITY CHECK CODE (QCC) 149. BASED ON THE INFO PROVIDED BY THE CUSTOMER, THERE WAS NO INDICATION OF PT IMPACT DUE TO INCREASE IN QCC 149S.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT CTNI CARTRIDGE | CTNI CARTRIDGE | MMI | ABBOTT POINT OF CARE CANADA LTD | S10130A |