FDA Adverse Event Malfunction Summary report: N

PROLENE SUT 30IN(75CM) 6-0 BLU

MDR report key: 19280343 · Received May 9, 2024

Report

Report Number
2210968-2024-05293
Event Type
Malfunction
Date Received
May 9, 2024
Date of Event
April 26, 2024
Report Date
June 27, 2024
Manufacturer
ETHICON INC.
Product Code
GAW
PMA / PMN Number
K133356
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ATTEMPTS ARE BEING MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. H6 COMPONENT CODE: G07002 ¿ DEVICE NOT RETURNED. RELATED EVENTS CAPTURED WITHIN: 2210968-2024-05294, 2210968-2024-05295 AND 2210968-2024-05296.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). DATE SENT TO THE FDA: 6/27/2024. H6 COMPONENT CODE: G07002 NO DEVICE PROBLEM FOUND. A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED, AND NO RELATED NON-CONFORMANCES WERE IDENTIFIED. H3 INVESTIGATIONAL SUMMARY: THE PRODUCT WAS RETURNED TO ETHICON FOR EVALUATION. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE CONDUCTED ON THE RETURNED DEVICE. THE RETURNED SAMPLE DETERMINED THAT IT WAS RECEIVED, TWENTY-EIGHT UNOPENED SAMPLES THAT PERTAIN TO THE PRODUCT CODE EH7228H. THE PRODUCT CODE IS DOUBLE-ARMED. AS PER THE SAMPLING PLAN, A VISUAL INSPECTION WAS PERFORMED ON THIRTEEN SAMPLES, THE PACKETS WERE OPENED. THE SWAGE AND ATTACHMENT AREA WERE NOTED TO BE AS EXPECTED. THE SUTURES WERE DISPENSED WITHOUT PROBLEMS AND EXAMINED ALONG THE STRAND NO ANOMALIES WERE OBSERVED DURING THE EVALUATION. THE FUNCTIONAL TEST WAS PERFORMED USING INSTRON EQUIPMENT AND THE PULL FORCE RESULT WAS ABOVE THE MINIMUM REQUIREMENTS. AS PART OF THE ETHICON QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT A VASCULAR PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE NEEDLE CAME LOOSE FROM THE SUTURE DURING VASCULAR ANASTOMOSIS. A NEW PACKAGE WAS TAKEN TO RESOLVE THE ISSUE. THERE WERE NO PATIENT CONSEQUENCES. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469974 PROLENE SUT 30IN(75CM) 6-0 BLU SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE GAW ETHICON INC. TKBJAU

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown