BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES
Report
- Report Number
- 1917413-2024-00392
- Event Type
- Malfunction
- Date Received
- May 8, 2024
- Date of Event
- April 11, 2024
- Report Date
- June 5, 2024
- Manufacturer
- BECTON, DICKINSON & CO. (BROKEN BOW)
- Product Code
- GIM
- UDI-DI
- 30382903678618
- PMA / PMN Number
- BK050036
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H6: INVESTIGATION SUMMARY: MATERIAL #: 367861. LOT/BATCH #: 3222735. BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 100 RETENTION SAMPLES FROM BD INVENTORY WERE EVALUATED BY VISUAL EXAMINATION AND NO ISSUES WERE OBSERVED RELATING TO CLOTTING AS ALL SAMPLES MET SPECIFICATIONS. COMPLAINTS FOR SAMPLE QUALITY ARE UNDER STATISTICAL CONTROL FOR THE MONTH OF APRIL 2024. AT THIS TIME, FURTHER TESTING IS NOT INDICATED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE CLOTTING. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. BD QUALITY WILL CONTINUE TO MONITOR SAMPLE QUALITY COMPLAINTS.
D.2. MEDICAL DEVICE TYPE: ONE ADDITIONAL CODE APPLIES, JKA G.5. PMA/510(K)#: ONE ADDITIONAL CODE APPLIES, K213670 H3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHILE USING BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES, MICRO-CLOTTING WAS FOUND IN AN UNSPECIFIED NUMBER OF SAMPLES INTERFERING WITH CBC TEST. SAMPLES NEEDED TO BE RECOLLECTED. NO IMPACT WAS REPORTED.
IT WAS REPORTED THAT WHILE USING BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES, MICRO-CLOTTING WAS FOUND IN AN UNSPECIFIED NUMBER OF SAMPLES INTERFERING WITH CBC TEST. SAMPLES NEEDED TO BE RECOLLECTED. NO IMPACT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 753539 | BD VACUTAINER® K2 EDTA (K2E) 7.2MG BLOOD COLLECTION TUBES | BLOOD SPECIMEN COLLECTION DEVICE | GIM | BECTON, DICKINSON & CO. (BROKEN BOW) | 3222735 | 30382903678618 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |