FDA Adverse Event Injury Summary report: N

ATTAIN STARFIX

MDR report key: 1926595 · Received December 14, 2010

Report

Report Number
2649622-2010-14420
Event Type
Injury
Date Received
December 14, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P060039
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND. THE FULL LEAD WAS RETURNED AND ANALYZED. THE DISTAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED); THERE WAS BLOOD/BODY FLUID ON OUTER TUBING OVERLAY; OUTER TUBING KINKED/BUCKLED; HELIX/LOBE DISTORTED/BENT; BLOOD IN/ON HELIX/LOBE MECHANISM. THERE WAS APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR LEAD WAS REMOVED AND REPLACED DUE TO PHRENIC NERVE STIMULATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN STARFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4195 ASKU

Patients

Seq Age Sex Outcome Treatment
1 56 YR Hospitalization| R D224TRK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6947 IMPLANTABLE TACHY LEAD| 5592 IMPLANTABLE PACING LEAD