SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2010-14414
- Event Type
- Malfunction
- Date Received
- December 14, 2010
- Date of Event
- October 15, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S039
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND THE HELIX DISENGAGED FROM THE HELICAL CHANNEL. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM (SLEEVE HEAD), THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE TIP SEAL WAS OBSERVED TO BE OUT OF POSITION.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
IT WAS REPORTED THAT DURING IMPLANT THE LEAD WAS BEING REPOSITIONED DUE TO A DILATED RIGHT VENTRICLE. ON THE THIRD ATTEMPT TO REPOSITION THE LEAD THE HELIX FAILED TO DEPLOY BECAUSE OF TISSUE ON THE TIP OF THE LEAD. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT DURING IMPLANT, THE LEAD WAS BEING REPOSITIONED DUE TO A DILATED RIGHT VENTRICLE. ON THE THIRD ATTEMPT TO REPOSITION THE LEAD, THE HELIX FAILED TO DEPLOY BECAUSE OF TISSUE ON THE TIP OF THE LEAD. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE S | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6935 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR | Other |