FDA Adverse Event
Injury
Summary report: N
SPRINT QUATTRO SECURE
MDR report key: 1926547
·
Received December 14, 2010
Report
- Report Number
- 2649622-2010-14389
- Event Type
- Injury
- Date Received
- December 14, 2010
- Date of Event
- October 19, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND; THE FULL LEAD WAS RETURNED AND ANALYZED. THE DISTAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED).
Description of Event or Problem · 1
IT WAS REPORTED THAT AFTER IMPLANT THE LEFT VENTRICULAR (LV) LEAD DISLODGED. DURING REPLACEMENT OF THE LV LEAD IT WAS NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS FOUND TO BE IN THE RIGHT ATRIUM. THE RV LEAD WAS REPOSITIONED AND THE LV LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD |