FDA Adverse Event
Injury
Summary report: N
INSYNC SENTRY
MDR report key: 1926344
·
Received December 14, 2010
Report
- Report Number
- 6000094-2010-02314
- Event Type
- Injury
- Date Received
- December 14, 2010
- Manufacturer
- MEDTRONIC S.A.
- Product Code
- LWS
- PMA / PMN Number
- ASKU
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED OR CONFIRMED MALFUNCTION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT EXPERIENCE SYNCOPAL EPISODES, THERE WAS HIGH AND VARYING IMPEDANCE ON THE RIGHT VENTRICULAR LEAD. THE SETSCREWS WERE ADJUSTED IN A REVISION. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INSYNC SENTRY | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC S.A. | 7298 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| L| R |