FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1926299 · Received December 14, 2010

Report

Report Number
2649622-2010-14272
Event Type
Injury
Date Received
December 14, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE CLINIC WITH A SHOCK. THE LEAD WAS OVERSENSING AND NOISE WAS NOTED ON ELECTROCARDIOGRAM . THE LEAD WAS THOUGHT TO HAVE AN APPARENT FRACTURE. THE LEAD WAS CAPPED DUE TO THE INABILITY TO RETRACT THE HELIX AND WAS REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 69 YR Hospitalization| L| R ICF09B IMPLANTABLE PACING LEAD| C154DWK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4195 IMPLANTABLE PACING LEAD