FDA Adverse Event
Injury
Summary report: N
ATTAIN ABILITY
MDR report key: 1926222
·
Received December 14, 2010
Report
- Report Number
- 2649622-2010-14238
- Event Type
- Injury
- Date Received
- December 14, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): NO ANOMALIES FOUND. THE FULL LEAD WAS RETURNED AND ANALYZED. (B)(4) NO ANOMALIES FOUND. THE FULL LEAD WAS RETURNED AND ANALYZED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEFT VENTRICULAR LEAD DISLODGED INTO THE RIGHT ATRIUM. IT WAS FURTHER REPORTED THAT DURING THE SUBSEQUENT LEFT VENTRICULAR LEAD REPLACEMENT PROCEDURE, THE INITIAL REPLACEMENT LEAD USED KEPT "SLIPPING BACK." BOTH LEADS WERE REMOVED AND REPLACED WITH A LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN ABILITY | IMPLANTABLE PACING LEAD | OJX | MEDTRONIC PUERTO RICO, INC. | 4196 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 88 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6947 IMPLANTABLE TACHY LEAD |