FDA Adverse Event Injury Summary report: N

VIRTUOSO VR

MDR report key: 1926126 · Received December 14, 2010

Report

Report Number
6000094-2010-02309
Event Type
Injury
Date Received
December 14, 2010
Manufacturer
MEDTRONIC S.A.
Product Code
LWS
PMA / PMN Number
ASKU
Removal / Correction Number
Z-0131-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) ANALYSIS OF THE DEVICE FOUND A HIGH CURRENT DRAIN CONDITION. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE MAY HAVE HAD EARLY BATTERY DEPLETION. DEVICE WAS EXPLANTED AND REPLACED. NO FUTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIRTUOSO VR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC S.A. D164VWC ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R