FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 1926124
·
Received November 24, 2010
Report
- Report Number
- 2531779-2010-02459
- Event Type
- Malfunction
- Date Received
- November 24, 2010
- Date of Event
- October 23, 2010
- Report Date
- October 23, 2010
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Removal / Correction Number
- 2531779-03/24/2010-003-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THERE IS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. ANIMAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. EVAL REVEALED A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED WHICH COULD NOT BE DUPLICATED DURING TESTING.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PISTON PUSHED ALL OF THE INSULIN OUT OF THE CARTRIDGE DURING THE LOAD CARTRIDGE PHASE. A FAMILY MEMBER STATED THAT THE PT WAS UNABLE TO USE THE PUMP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |