FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1926112 · Received November 24, 2010

Report

Report Number
2531779-2010-02476
Event Type
Malfunction
Date Received
November 24, 2010
Date of Event
October 24, 2010
Report Date
October 24, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Removal / Correction Number
2531779-03/24/2010-003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE USER GUIDE INSTRUCTS USERS TO DISCONNECT FROM THE INFUSION SET PRIOR TO INITIATING THE PRIME / REWIND SEQUENCE. ADDITIONALLY, THE PUMP NOTIFIES THE USER TO DISCONNECT FROM THE INFUSION SET PRIOR TO INITIATING THE REWIND OF THE MOTOR. THE USER MUST MANUALLY CONFIRM THIS ALERT BEFORE PROCEEDING WITH THE REWIND PROCESS. THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVAL SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION (B)(6) 2011 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP HISTORY INDICATED THAT THE PUMP WAS IN USE UNTIL (B)(6) 2011. THE DATE OF THE REPORTED EVENT WAS (B)(6) 2010. DUE TO CONTINUED PUMP USE, THE HISTORY FOR THE DATE OF THE REPORTED EVENT WAS UNAVAILABLE FOR REVIEW. DURING TESTING, REWIND, LOAD, AND PRIME STEPS WERE PERFORMED SUCCESSFULLY WITH NO ALARMS OBSERVED. EVALUATION CONFIRMED THAT THE PUMP DISPLAYS THE APPROPRIATE WARNINGS FOR THE USER TO DISCONNECT FROM THE PUMP BEFORE PERFORMING THE REWIND, LOAD, AND PRIME STEPS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THERE WAS NO DEFECT FOUND ON INVESTIGATION.

Description of Event or Problem · 1

THE PT'S MOTHER REPORTED THAT SHE FILLED THE CARTRIDGE TO THE 1ML (100UNIT) MARK AND COMPLETED THE REWIND STEP WITH THE PT DISCONNECTED FROM THE PUMP. SHE THEN CONNECTED THE PT TO THE PUMP AND COMPLETED THE LOAD CARTRIDGE STEP. THE PUMP READ 68 UNITS FOLLOWING THE LOAD CARTRIDGE STEP. THE PT'S MOTHER BELIEVES THAT INSULIN WAS INADVERTENTLY INJECTED INTO THE PT DURING THE LOAD CARTRIDGE PHASE. NO ADVERSE EVENT OR BLOOD GLUCOSE EXCURSIONS WERE REPORTED AS A RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 5 YR