FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1926065 · Received November 24, 2010

Report

Report Number
2531779-2010-02449
Event Type
Malfunction
Date Received
November 24, 2010
Date of Event
October 22, 2010
Report Date
October 22, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS AND THE EVAL REVEALED THAT THE KEYPAD APPEARED TO BE INTACT WITH NO LIFTING OR PEELING. ALL OF THE KEYPAD BUTTONS WERE INTERMITTENTLY RESPONDING AND REQUIRED EXCESSIVE FORCE TO ACTIVATE. THE KEYPAD WAS REMOVED AND THE "DOWN" AND "OK" CONTACTS WERE MISALIGNED. ALL OF THE KEY CONTACTS WERE INVERTED AND DID NOT ALWAYS SPRING BACK. THERE WAS ALSO EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

THE PT'S MOTHER REPORTED THAT THE KEYPAD WAS INTERMITTENTLY RESPONDING. THE PT REPORTED THAT THE "OK" BUTTON WAS SLOW TO RESPOND AND HAD TO BE PRESSED MULTIPLE TIMES TO ELICIT A RESPONSE FROM THE KEYPAD. THE OTHER BUTTONS WERE RESPONDING BUT SLOWER THAN USUAL. THE RPTR DENIES DAMAGE TO THE KEYPAD OR EXPOSURE TO MOISTURE AND CLEANING SOLVENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 15 YR