FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 1926026 · Received November 24, 2010

Report

Report Number
2531779-2010-02455
Event Type
Malfunction
Date Received
November 24, 2010
Date of Event
October 22, 2010
Report Date
October 22, 2010
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS AND THE KEYPAD APPEARED TO BE INTACT WITH NO LIFTING OR PEELING. ALL OF THE KEYPAD BUTTONS WERE INTERMITTENTLY RESPONDING AND REQUIRED EXCESSIVE FORCE TO ACTIVATE. THE KEYPAD WAS REMOVED AND THERE WAS EVIDENCE OF KEYPAD ADHESIVE UNDER KEY CONTACTS.

Description of Event or Problem · 1

THE PT REPORTED THAT THE KEYPAD BUTTONS WERE NOT RESPONDING CORRECTLY. THE NUMBERS WOULD NOT SCROLL, BUT WHEN THE DESIRED BOLUS AMOUNT WAS PROGRAMMED, THEY WOULD APPEAR FOR A SHORT TIME. THE NUMBERS WOULD THEN CONTINUE TO EITHER INCREASE OR DECREASE EVEN AFTER THE BUTTONS WERE NOT PRESSED DEPENDING ON WHICH ARROW BUTTONS WERE PRESSED CAUSING THE INCORRECT BOLUS AMOUNT. THE RPTR DENIES DAMAGE TO THE KEYPAD AND THE BUTTONS DID NOT STICK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONE TOUCH PING NA

Patients

Seq Age Sex Outcome Treatment
1 51 YR