SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2010-14121
- Event Type
- Injury
- Date Received
- December 14, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) THE DEVICE WAS ANALYZED AND THERE WERE NO ANOMALIES FOUND. EVALUATION SUMMARY: (B)(4): PARTS OF THE LEAD WERE RETURNED IN SEGMENTS AND ANALYZED. ANALYSIS RESULTS REVEALED THAT NO ANOMALIES WERE FOUND. BLOOD/BODY FLUID WAS PRESENT ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED) AND IN/ON THE HELIX MECHANISM. IT WAS ALSO NOTED THAT THE INNER TUBING WAS KINKED/BUCKLED, THE OUTER TUBING OVERLAY WAS MELTED, THERE WAS A COSMETIC DEPRESSION ON THE OUTER INSULATION, AND THERE WAS APPARENT EXPLANT DAMAGE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) THE DEVICE WAS ANALYZED AND THERE WERE NO ANOMALIES FOUND. (B)(4) ANALYSIS OF THE LEAD IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
IT WAS REPORTED THAT THE DEVICE AND RIGHT VENTRICULAR LEAD WERE REMOVED AT THE PATIENT REQUEST BECAUSE IT WAS HURTING. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Hospitalization| R |