FDA Adverse Event Malfunction Summary report: N

PLUM XL3 PLS MIC/MAC

MDR report key: 1925930 · Received November 8, 2010

Report

Report Number
2921482-2010-00866
Event Type
Malfunction
Date Received
November 8, 2010
Date of Event
September 20, 2010
Report Date
October 13, 2010
Manufacturer
HOSPIRA, INC.
Product Code
FRN
PMA / PMN Number
K952799
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE WAS ABLE TO BE PROGRAMMED WHILE THE LOCKOUT SWITCH WAS IN THE LOCKED POSITION. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PLUM XL3 PLS MIC/MAC 80FRN FRN HOSPIRA, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA