FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1925927 · Received December 14, 2010

Report

Report Number
2649622-2010-14103
Event Type
Malfunction
Date Received
December 14, 2010
Date of Event
October 18, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE LEAD IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS RESULTS REVEALED THAT NO ANOMALIES WERE FOUND. BLOOD WAS PRESENT IN/ON THE HELIX MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED DURING IMPLANT ATTEMPT OF THE NEW LEAD, THE LEAD WOULD NOT PASS THE SVC COIL AS THE ANATOMY WAS TOO TIGHT. THE LEAD WAS NOT USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED DURING IMPLANT ATTEMPT OF THE NEW LEAD, THE LEAD WOULD NOT PASS THE SUPERIOR VENA CAVA COIL AS THE ANATOMY WAS TOO TIGHT. THE LEAD WAS NOT USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 82 YR Other 4568 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD| 4194 IMPLANTABLE PACING LEAD| 4568 IMPLANTABLE PACING LEAD