SPRINT
Report
- Report Number
- 2649622-2010-14101
- Event Type
- Malfunction
- Date Received
- December 14, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S13
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. ANALYSIS RESULTS REVEALED THE PRESENCE OF OVERSENSING. THERE WERE NINE VENTRICULAR NON-SUSTAINED TACHYCARDIA EPISODES WITH LESS THAN 200 MS AVERAGE VENTRICULAR CYCLE LENGTH RECORDED BETWEEN (B)(6) 2010 AND (B)(6) 2010. ANALYSIS ALSO NOTED INTERFERENCE/NOISE WAS RECORDED. THE VENTRICULAR SHORT INTERVAL COUNT AVERAGED 2977.5 COUNTS/DAY, OVER 35 DAYS, BETWEEN (B)(6) 2010 AND (B)(6) 2010.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THE DEVICE SHOWS A HIGH NUMBER OF SHORT INTERVAL COUNTS AND OVERSENSING WAS NOTED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6943 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Other | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5568 IMPLANTABLE PACING LEAD| 5568 IMPLANTABLE PACING LEAD |