FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1925916 · Received December 14, 2010

Report

Report Number
2649622-2010-14099
Event Type
Injury
Date Received
December 14, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT CAME TO THE HOSPITAL BECAUSE OF RECEIVING SHOCKS DUE TO T-WAVE OVERSENSING AND DECREASED R-WAVE SENSING. THE DEFIBRILLATOR WAS EXPLANTED AND REPLACED. THE PACE/SENSE PORTION OF THE RIGHT VENTRICULAR LEAD WAS CAPPED, AND A NEW PACE/SENSE LEAD WAS IMPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT CAME TO THE HOSPITAL BECAUSE OF RECEIVING SHOCKS DUE TO T-WAVE OVERSENSING. THE DEFIBRILLATOR WAS EXPLANTED AND REPLACED. THE RIGHT VENTRICULAR PACE/SENSE LEAD IS STILL IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R 6949 IMPLANTABLE TACHY LEAD