FDA Adverse Event Malfunction Summary report: N

I-STAT ACT KAOLIN CARTRIDGE

MDR report key: 19258849 · Received May 7, 2024

Report

Report Number
2245578-2024-00087
Event Type
Malfunction
Date Received
May 7, 2024
Date of Event
April 29, 2024
Report Date
July 12, 2024
Manufacturer
ABBOTT POINT OF CARE INC.
Product Code
JBP
PMA / PMN Number
K023582
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 09-JUL-2024. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT PASSED RELEASE SPECIFICATIONS. RETAINED AND RETURNED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA IN APPENDIX 1 OF Q04.01.003 REV. AN (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN IDENTIFIED.

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 0

ON 30-APR-2024, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT ACT KAOLIN CARTRIDGES THAT YIELDED A DISCREPANT RESULTS ON A 62 YEAR OLD MALE WITH ABNORMAL DOBUTAMINE STRESS ECHO. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD DATE TESTED ACT(SEC) I-STAT (B)(6) 2024 14:12 509 I-STAT (B)(6) 2024 14:22 255 I-STAT (B)(6) 2024 14:38 514 I-STAT (B)(6) 2024 14:49 282 PER I-STAT SYSTEM MANUAL (ART: 714185-00Q), THE I-STAT KAOLIN ACTIVATED CLOTTING TIME (KAOLINACT) TEST IS AN IN VITRO DIAGNOSTIC TEST THAT USES FRESH, WHOLE BLOOD. AND IS USED TO MONITOR HIGH-DOSE HEPARIN ANTICOAGULATION FREQUENTLY ASSOCIATED WITH CARDIOVASCULAR SURGERY. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. AN INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
641842 I-STAT ACT KAOLIN CARTRIDGE ACT KAOLIN CARTRIDGE JBP ABBOTT POINT OF CARE INC. NA R24056

Patients

Seq Age Sex Outcome Treatment
1 62 YR Male