PTERYFIT IMPLANT
Report
- Report Number
- 3011390931-2024-00062
- Event Type
- Injury
- Date Received
- May 7, 2024
- Date of Event
- January 24, 2024
- Report Date
- May 7, 2024
- Manufacturer
- NORIS MEDICAL LTD
- Product Code
- DZE
- UDI-DI
- 07290108817133
- PMA / PMN Number
- K140440
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
THE DHR WAS REVIEWED FOR LOT#9000395 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE INVESTIGATION FOR THIS EVENT HAS NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. OUR MANUFACTURING SYSTEM ASSURES, THAT PRODUCTION AND PROCESS CONTROLS ARE IN PLACE TO ENSURE THAT BATCHES CONFORM TO THE APPLICABLE SPECIFICATIONS BEFORE THEY ARE DISTRIBUTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.
UDI RELATED DATA QUALITY UPDATES ONLY.
THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2024. ON (B)(6) 2024, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT HAS HYPERTENSION . OBSERVED DURING THE EVENT: MOBILITY. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 454169 | PTERYFIT IMPLANT | PTERYFIT DENTAL IMPLANT | DZE | NORIS MEDICAL LTD | NMAF4225 | 9000395 | 07290108817133 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Male |