FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1925849 · Received December 14, 2010

Report

Report Number
2649622-2010-14073
Event Type
Injury
Date Received
December 14, 2010
Date of Event
September 15, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND THE HELIX DISENGAGED FROM THE HELICAL CHANNEL. IT WAS NOTED THAT THE INNER TUBING WAS KINKED/BUCKLED, THE CONNECTOR PIN BENT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM (SLEEVE HEAD), THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE TIP SEAL WAS OBSERVED TO BE OUT OF POSITION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS OVERSENSING, NOISE, HIGH IMPEDANCE, AND LEAD INTEGRITY ALERT WAS TRIGGERED ON THE RIGHT VENTRICULAR LEAD DUE TO A SUSPECTED INSULATION BREACH. IT WAS ALSO REPORTED THAT A VENTRICULAR TACHYCARDIA EPISODE COULD NOT BE TERMINATED WITH THE DELIVERED ANTI-TACHYCARDIA PACING AND CARDIOVERSION THERAPIES. THE LEAD WAS EXTRACTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 54 YR Hospitalization| R