PUMP MMT-1884L MM780G BLE 3.0 MG/ML
Report
- Report Number
- 2032227-2024-164932
- Event Type
- Malfunction
- Date Received
- May 7, 2024
- Date of Event
- April 15, 2024
- Report Date
- June 6, 2024
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- PMA / PMN Number
- P160017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- 003
Narratives
THE PUMP PASSED THE DISPLACEMENT TEST, ACTIVE CURRENT MEASUREMENT TEST, SLEEP CURRENT MEASUREMENT TEST AND SELF-TEST. PUMP ERROR 23 WAS NOTED WHICH WAS EXPECTED. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUMP. THE POWER MANAGEMENT GRAPH CONFIRMED THE UNLOADED VOLTAGE (UL VLITH) AND LOADED VOLTAGE (LOADED VLITH) WERE WITHIN SPEC RANGE. PUMP ERROR 23 WAS FOUND IN THE HISTORY FILES ON (B)(6) 2024 23:23:52.000. PUMP ERROR 49 WAS FOUND IN THE HISTORY FILES ON (B)(6) 2024 23:19:12.000. PUMP ERROR 68 WAS FOUND IN THE HISTORY FILES ON (B)(6) 2024 23:19:12.000. P-CAP LOCKS PROPERLY INTO THE RESERVOIR COMPARTMENT. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND MOISTURE DAMAGE TO THE PCBA 1, PCBA 2 AND FORCE SENSOR. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: PILLOWING KEYPAD OVERLAY, STAINED KEYPAD OVERLAY, CRACKED KEYPAD OVERLAY, SCRATCHED CASE, CRACKED CASE, CRACKED CASE (BATTERY TUBE), BATTERY TUBE THREADS - CRACKED AND LABEL DAMAGE (FADED). PUMP ERROR 23, 49 AND 68 WERE NOT CONFIRMED. COSMETIC DAMAGE CONFIRMED AT BATTERY SIDE. EXPOSED TO MOISTURE CONFIRMED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED A CRACKED PUMP AND RECEIVED PUMP ERROR 49,ERROR 68 AND ERROR 23. THE CUSTOMER REPORTED NO ADVERSE EVENT. TROUBLESHOOTING WAS PERFORMED. THE EVENT INVOLVED PRODUCT(S) MMT-1884L. CUSTOMER REPORTED PHYSICAL DAMAGE (CRACK OR SCRATCH) ON THE PUMP NOT RELATED TO THE RETAINER RING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. MMT-1884L WAS REQUESTED AND CUSTOMER RESPONSE WAS THE DEVICE WILL BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 642833 | PUMP MMT-1884L MM780G BLE 3.0 MG/ML | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1884L | NG3202001H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Male |