SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2010-13942
- Event Type
- Malfunction
- Date Received
- December 14, 2010
- Date of Event
- October 12, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS RESULTS REVEALED THAT THE HELIX WAS DISENGAGED FROM THE HELICAL CHANNEL. BLOOD/BODY FLUID WAS PRESENT ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED) AND IN/ON THE HELIX MECHANISM AND ITS SLEEVE HEAD. THE TIP SEAL WAS OBSERVED TO BE OUT OF POSITION.
IT WAS REPORTED THAT DURING THE ATTEMPTED IMPLANT OF THE RIGHT VENTRICULAR LEAD, THE HELIX WOULD NOT DEPLOY FROM THE LEAD BODY. MULTIPLE ATTEMPTS ["25+ TURNS"] FAILED. THE LEAD WAS NOT IMPLANTED AND A NEW LEAD WAS PLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Other |