FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1925580 · Received December 14, 2010

Report

Report Number
2649622-2010-13942
Event Type
Malfunction
Date Received
December 14, 2010
Date of Event
October 12, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS RESULTS REVEALED THAT THE HELIX WAS DISENGAGED FROM THE HELICAL CHANNEL. BLOOD/BODY FLUID WAS PRESENT ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED) AND IN/ON THE HELIX MECHANISM AND ITS SLEEVE HEAD. THE TIP SEAL WAS OBSERVED TO BE OUT OF POSITION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE ATTEMPTED IMPLANT OF THE RIGHT VENTRICULAR LEAD, THE HELIX WOULD NOT DEPLOY FROM THE LEAD BODY. MULTIPLE ATTEMPTS ["25+ TURNS"] FAILED. THE LEAD WAS NOT IMPLANTED AND A NEW LEAD WAS PLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 60 YR Other